|
### product event summary: three (3) batteries ((b)(6)) were returned for evaluation.No performance allegations were made against the batteries.Failure analysis of the returned batteries revealed that all the devices passed visual inspection.Functional testing of the batteries revealed abnormalities relating to the relative state of charge (rsoc); the batteries were not estimating the capacity accurately, likely due to an estimation error.Furthermore, it was observed that the batteries had exceeded its useful operating life of 500 charge and discharge cycles.An internal visual inspection of the batteries did not reveal any anomalies.Additional products: d1: heartware ventricular assist system ¿ battery d4: model #: 1650 / catalog #: 1650 / expiration date: 31-jan-2020 / serial #: (b)(6) udi #: (b)(4).D9: yes, return date: 22-jun-2023.H3: yes dev rtn to mfr? yes.H4: mfg date: 17-jan-2019.H5: no.D1: heartware ventricular assist system ¿ battery d4: model #: 1650 / catalog #: 1650 / expiration date: 31-jan-2020 / serial #: (b)(6) udi #: (b)(4).D9: yes, return date: 22-jun-2023.H3: yes dev rtn to mfr? yes.H4: mfg date: 14-jan-2019.H5: no.H6: the codes present in section h6 correspond to components/products that comprise the reported event.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|