| Catalog Number |
100008633 |
| Medical Device Problem Code |
Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.
|
| |
|
Event or Problem Description
|
|
While using a schick 33 sensor intraoral x-ray unit, an incorrect image was displayed during use.This was immediately noticeable to the clinician and no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device that was reported initially for this event is being corrected from schick 33 with ae usb to schick33 s2 sensor starter kit/3.0 interface 6ft.
|
| |
|