| Model Number |
106130-17 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Skin Inflammation/ Irritation (4545)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported by the patient that her skin became very red and itchy while using 72r electrodes and cover patches.The patient changed them every day and rotated at that time.She wears them on her lower back.The patient is not sure if the irritation was coming from the tape that was on her wound.She spoke with her doctor, and he gave her medication for the itching.The patient states she does have sensitive skin.The customer service representative advised the patient to pause treatment until her skin clears.The patient was sent replacement 63b electrodes.No further consequences have been reported at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported by the patient that her skin became very red and itchy while using 72r electrodes and cover patches.The patient changed them every day and rotated at that time.She wears them on her lower back.The patient is not sure if the irritation was coming from the tape that was on her wound.She spoke with her doctor, and he gave her medication for the itching.The patient states she does have sensitive skin.The customer service representative advised the patient to pause treatment until her skin clears.The patient was sent replacement 63b electrodes.No further consequences have been reported at this time.The product was not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information and corrected information based on udi related data quality updates.The device was not returned to ebi for evaluation.The reported event was unable to be confirmed due to limited information received from the customer.The device history record was reviewed, and no discrepancies related to the reported event were found.The root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Corrections: d1: brand name, d2a: device common name,d4: model number, g4: pma number, h6 product and evaluation codes.Additional information: a: patient information, e: initial reporter.
|
| |
|
Event or Problem Description
|
|
It was reported by the patient that her skin became very red and itchy while using 72r electrodes and cover patches.The patient changed them every day and rotated at that time.She wears them on her lower back.The patient is not sure if the irritation was coming from the tape that was on her wound.She spoke with her doctor, and he gave her medication for the itching.The patient states she does have sensitive skin.The customer service representative advised the patient to pause treatment until her skin clears.The patient was sent replacement 63b electrodes.No further consequences have been reported at this time.The product was not returned.
|
| |
|