| Model Number |
27040DB |
| Medical Device Problem Codes |
Energy Output Problem (1431); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
08/25/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that there was an issue with a working element.There is no failure description available.According to the information received there was an additional medical intervention necessary due to the incident reported and a poor haemostasis occurred.Due to limited information and the information mentioned above this case was rated deemed reportable.Additional information was already requested.If additional information becomes available this report will be updated unsolicited.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in section d10 to add the associated product.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was further reported that the defective bipolar electric cable had smoke coming out of the handle holder and intermediate part of the device was overheating.Hemostasis was less effective, with bleeding persisting despite theoretically effective coagulation.Hematuria occurred.Bleeding was stopped using a monopolar resector instead.There is no permanent or temporary impairment for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in section b5.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
According to the information received from the customer, there was poor hemostasis and poor cutting quality during the operation with the items.In addition, the customer reported that between the article 27040db working element, bipolar and the connected bipolar electric cable (manufacturer, article number unknown) there was an overheating and smoke development from the cable.It is assumed that due to residual moisture or moisture intrusion a short circuit occurred between the cable connector and the working element.The short circuit may have led to the described overheating and smoke development.The moisture intrusion is assumed to be caused due to improper handling.A more detailed investigation cannot be carried out without returned articles.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|