• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG RIGID ENDOSCOPE WORKING GUIDE; WORKING ELEMENT, BIPOLAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG RIGID ENDOSCOPE WORKING GUIDE; WORKING ELEMENT, BIPOLAR Back to Search Results
Model Number 27040DB
Medical Device Problem Codes Energy Output Problem (1431); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 08/25/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that there was an issue with a working element.There is no failure description available.According to the information received there was an additional medical intervention necessary due to the incident reported and a poor haemostasis occurred.Due to limited information and the information mentioned above this case was rated deemed reportable.Additional information was already requested.If additional information becomes available this report will be updated unsolicited.
 
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
 
Additional Manufacturer Narrative
Additional information is provided in section d10 to add the associated product.The event is filed under internal karl storz complaint id: (b)(4).
 
Event or Problem Description
It was further reported that the defective bipolar electric cable had smoke coming out of the handle holder and intermediate part of the device was overheating.Hemostasis was less effective, with bleeding persisting despite theoretically effective coagulation.Hematuria occurred.Bleeding was stopped using a monopolar resector instead.There is no permanent or temporary impairment for the patient.
 
Additional Manufacturer Narrative
Additional information is provided in section b5.The event is filed under internal karl storz complaint id: (b)(4).
 
Additional Manufacturer Narrative
According to the information received from the customer, there was poor hemostasis and poor cutting quality during the operation with the items.In addition, the customer reported that between the article 27040db working element, bipolar and the connected bipolar electric cable (manufacturer, article number unknown) there was an overheating and smoke development from the cable.It is assumed that due to residual moisture or moisture intrusion a short circuit occurred between the cable connector and the working element.The short circuit may have led to the described overheating and smoke development.The moisture intrusion is assumed to be caused due to improper handling.A more detailed investigation cannot be carried out without returned articles.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIGID ENDOSCOPE WORKING GUIDE
Common Device Name
WORKING ELEMENT, BIPOLAR
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key17738497
Report Number9610617-2023-00255
Device Sequence Number6634368
Product Code FAS
UDI-Device Identifier04048551075500
UDI-Public4048551075500
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K221893
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number27040DB
Device Catalogue Number27040DB
Device Lot NumberXR114
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2023
Supplement Date Received by Manufacturer08/28/2023
10/20/2023
12/18/2023
Initial Report FDA Received Date09/13/2023
Supplement Report FDA Received Date09/21/2023
10/31/2023
12/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UH300 - AUTOCON III 300,220-240V.
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-