LEXINGTON MEDICAL, INC. AEON ENDOSCOPIC STAPLER; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number AESH160
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2023
Type of Reportable Event Malfunction
Event or Problem Description
An aeon endostapler handle medium was used with a 45 purple reload during a lobectomy procedure on (b)(6) 2023.During the procedure the aesr45p was fired on the bronchus, after the firing the reload would not unclamp and retract from the tissue.The surgeon needed to use a scalpel to release the reload and suture the bronchus.No patient harm.During a follow-up conversation, the surgeon stated that the tissue had been previously irradiated and was significantly inflamed.They also said the tissue was thick and in hindsight, a 45mm black reload may have been more appropriate for the tissue.
 
Additional Manufacturer Narrative
The device was returned and investigated no defects were found in the returned product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AEON ENDOSCOPIC STAPLER
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
LEXINGTON MEDICAL, INC.
23 crosby drive
bedford MA 01730
Manufacturer (Section G)
LEXINGTON MEDICAL INC.
23 crosby drive
bedford MA 01730
Manufacturer Contact
rainer maas
23 crosby drive
bedford, MA 01730
6172099817
MDR Report Key17741105
Report Number3012998149-2023-00005
Device Sequence Number6615349
Product Code GDW
UDI-Device Identifier00851574008004
UDI-Public(01)00851574008004(17)280229(10)23BH128
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
222210
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberAESH160
Device Lot Number23BH128
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2023
Initial Report FDA Received Date09/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AESR45P AEON¿ ENDOSCOPIC STAPLER 4.0MM/45MM RELOAD
Outcome Attributed to Adverse Event Other;
Patient Age53 YR
Patient SexFemale
-
-