Model Number VICM512.6
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2023
Type of Reportable Event Malfunction
Event or Problem Description
The reporter indicated that 12.6mm vicm512.6 implantable collamer lens of a -7.0 diopter tore/broke during injection into the patient's right eye (od) on (b)(6) 2023.There was patient contact, but no patient injury.Cause of the event was user error, with the device failing to perform.Reportedly, "the lens foot plate was missing and likely it was caught on the cartridge and was only noticed when it was in the eye already." a replacement lens was implanted and the problem was resolved.
 
Additional Manufacturer Narrative
A4-a6:unk.H6: user error.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
800 royal oaks suite:210
monrovia, CA 91016
6263037902
MDR Report Key17741570
Report Number2023826-2023-03992
Device Sequence Number4235995
Product Code MTA
UDI-Device Identifier00841542119122
UDI-Public00841542119122
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model NumberVICM512.6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2023
Initial Report FDA Received Date09/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: SFC-45 - LOT#UNK.; FOAM TIP PLUNGER - LOT# UNK.; INJECTOR MODEL: LIOLI-24 - LOT# FCK2004.
Patient Age26 YR
Patient SexFemale
-
-