| Model Number |
M00517400 |
| Medical Device Problem Codes |
Premature Activation (1484); Material Deformation (2976); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Block e1: the reported healthcare facility is: (b)(6) medical center.Block h6: imdrf device code a1502 captures the reportable event of stent positioning issue.Imdrf device code a0406 captures the reportable event of stent material deformation.Imdrf device code a150103 captures the reportable event of stent prematurely deployed.Imdrf impact code f2303 captures the reportable event of administering steroids.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an agile esophageal stent was to be implanted to treat an anastomatic stricture in the esophagus during an esophageal stent placement procedure performed on (b)(6) 2023.During the procedure, the stent prematurely deployed before the distal handle was completely pulled back, which caused the stent to become deformed and migrate into the patient's stomach.The stent was removed, and the physician completed the procedure by administering steroids and using a cre balloon.The patient will be rescheduled for another stent placement procedure.There were no patient complications as a result of this event.
|
| |
|
Search Alerts/Recalls
|