• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Model Number 24657
Medical Device Problem Codes Premature Activation (1484); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the stent inadvertently deployed.This 7x60, 130 cm eluvia drug-eluting vascular stent system was selected for use in a runoff procedure for peripheral artery disease.The stent system was loaded over a non-boston scientific guidewire.As soon as the stent was loaded on the wire, it was difficult to advance; therefore, the stent system was removed from the wire.However, when pulling it off the wire the stent inadvertently deployed.It was noted that the stent never entered the patients body.The procedure was completed with a new stent and there were no patient complications.
 
Event or Problem Description
It was reported that the stent inadvertently deployed.This 7x60, 130 cm eluvia drug-eluting vascular stent system was selected for use in a runoff procedure for peripheral artery disease.The stent system was loaded over a non-boston scientific guidewire.As soon as the stent was loaded on the wire, it was difficult to advance; therefore, the stent system was removed from the wire.However, when pulling it off the wire the stent inadvertently deployed.It was noted that the stent never entered the patients body.The procedure was completed with a new stent and there were no patient complications.
 
Additional Manufacturer Narrative
Device eval by manufacturer: upon receipt at our post market quality assurance laboratory, this eluvia drug-eluting vascular stent system was inspected for damage.Visual examination revealed a kink to the outer sheath at the nosecone.The stent was partially deployed 15mm from the distal end of the middle sheath.Microscopic examination revealed no additional damages.The thumbwheel lock and pull rack were still in the manufactured position.Inspection of the remainder of the device, revealed no other damage or irregularities.Product analysis confirmed that the stent was partially deployed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17745131
Report Number2124215-2023-50277
Device Sequence Number5654270
Product Code NIU
UDI-Device Identifier08714729876700
UDI-Public08714729876700
Combination Product (Y/N)Y
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P180011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/06/2025
Device Model Number24657
Device Catalogue Number24657
Device Lot Number0031634909
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/12/2023
Supplement Date Received by Manufacturer10/11/2023
Initial Report FDA Received Date09/14/2023
Supplement Report FDA Received Date10/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GLIDE ADVANTAGE WIRE; GLIDE ADVANTAGE WIRE
Patient SexUnknown
-
-