| Model Number |
35615 |
| Medical Device Problem Codes |
Defective Device (2588); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the ro marker was pushed.The target lesion was located in the biliary.A accustick ii w/o guidewire was selected for use.During the procedure, the dilator was inserted into the patient, but could not see the angiomark.The device was removed from the patient and found that the angiomark was pushed.The procedure was completed with another of the same device.No complications were reported and the patient was fine.
|
| |
|
Event or Problem Description
|
|
It was reported that the ro marker was pushed.The target lesion was located in the biliary.A accustick ii w/o guidiwire was selected for use.During the procedure, the dilator was inserted into the patient, but could not see the angiomark.The device was removed from the patient and found that the angiomark was pushed.The procedure was completed with another of the same device.No complications were reported, and the patient was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the accustick device was returned with the dilator and the canula stuck inside the device.It was observed that the ro marker was out of its location due it was separated from the accustick device.Additionally, the accustick device was kinked and had the distal tip was damage.No more damages were observed.The device was inspected under microscope, and it was observed that the distal tip was damage, this damage is located next to the ro marker position.It's important to mention that evidence of correct ro marker colocation was observed at device distal end section.
|
| |
|
Search Alerts/Recalls
|