ZIMMER MANUFACTURING B.V. 16MM DIAMETER 100MM LENGTH STRAIGHT STEM EXTENSION COMBINED LENGTH 145MM; PROSTHESIS, KNEE Back to Search Results
Catalog Number 00598801016
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Joint Dislocation (2374); Ambulation or Postural Difficulties (2544); Joint effusion (4849)
Date of Event 07/18/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent a revision total knee arthroplasty.Approximately one year and nine months post-implantation, the patient began to experience pain, instability, immobility and moderate joint effusion.Diagnostic images revealed that the locking bar component of the articular surface had fractured.Subsequently, the patient underwent revision surgery.During the surgery, the knee joint was found to be uncoupled from the connector due to breakage of the locking bolt as well as dorsal dislocation of the articular surface and loosening of the cemented femoral component.The femoral components and articular surface were revised without complication.Due diligence is complete as multiple attempts have been made; all available information has been reported.
 
Additional Manufacturer Narrative
(b)(4).D10: medical product: 12mm height size e articular surface: catalog#00588005012, lot#64834891; left size e cemented option femoral component: catalog#00588001501, lot#64366826; size 4 precoat cemented tibial component: catalog#00588000400, lot#64739910; 14mm diameter 100mm length straight stem extension combined length 145mm: catalog#00598801014, lot#64710790; trabecular metal tibial cone, medium 36x31mm: catalog#00545001336, lot#64819407; trabecular metal femoral diaphyseal cone, 30mm, small, left: catalog#00545001730, lot#64532796; trabecular metal femoral metaphyseal cone, 35mm, medium, left: catalog#00545002135, lot#63858698; mathys gelatin cement restrictor c-plug size 14: catalog#237014, lot#20f0905009, qty#2; refobacin plus bone cem 2x40-3: catalog#3021170001-3, lot#y01bal0104.G2: foreign: germany.Multiple mdr reports have been filed for this event.Please see associated reports: 0001822565-2023-01883; 0001822565-2023-02511; 0001822565-2023-02512; 0001822565-2023-02513.The product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Upon receipt of additional information or completion of the investigation, a follow-up mdr will be submitted.H3 other text : see h10 narrative.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Proposed component code: mechanical (g04) - stem visual and dimensional evaluation of the returned device found it to exhibit signs of being implanted (gouged/nicked/wear) and is covered in bone cement.Device history record was reviewed and no discrepancies related to the reported event were found.Medical records and radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: fracture, loosening, breakage of the hinge bolt, pain and instability noted.Radiographs show possible loosening.Root cause was unable to be determined.This complaint was confirmed for loosening and instability by the medical records.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information at time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
16MM DIAMETER 100MM LENGTH STRAIGHT STEM EXTENSION COMBINED LENGTH 145MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17745661
Report Number0002648920-2023-00218
Device Sequence Number5298244
Product Code JWH
UDI-Device Identifier00889024221000
UDI-Public(01)00889024221000(17)300331(10)64721298
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K173057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00598801016
Device Lot Number64721298
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2023
Supplement Date Received by Manufacturer12/11/2023
Initial Report FDA Received Date09/14/2023
Supplement Report FDA Received Date12/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight98 KG
-
-