• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1650
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/21/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A supplemental report is being submitted for analysis and investigation completion.Product event summary: the battery (b)(6) was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Log file analysis did not reveal any anomalies.Failure analysis of the battery revealed that the device passed visual inspection.As received, all four (4) battery display light emitting diodes (leds) on the battery were able to illuminate.Functional testing revealed that the battery was unable to provide power or charge due to an enabled zero-volt charge (zvchg) field-effect transistor (fet).The gas gauge is not programmed to use the zvchg fet, which indicates that the battery unintentionally enabled the fet.The zvchg fet disables the charge and discharge fets, rendering the battery inoperable.As a result, the reported display error event could not be confirmed; however, it is possible that the battery's inability to charge or provide power was perceived as the reported display error event.An internal inspection did not reveal any anomalies.(b)(6) is in scope of fa1265, which was initiated for batteries with electrical faults.Based on an investigation conducted under capa pr00519785, the most likely root cause of the observed inability of the battery to charge or provide power has been attributed to the battery connector interface design which does not guarantee the desired connection angle in some use conditions.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The battery was returned to the manufacturer due to a display error.The battery subsequently tested out of specification during manufacturer¿s analysis.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key17746028
Report Number3007042319-2023-03535
Device Sequence Number17202506
Product Code DSQ
UDI-Device Identifier00888707009942
UDI-Public00888707009942
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/30/2023
Device Model Number1650
Device Catalogue Number1650
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2023
Initial Report FDA Received Date09/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/07/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1512-2022
Patient Sequence Number1
Patient Age37 YR
Patient SexMale
Patient Weight78 KG
-
-