| Model Number |
DISPS KIT,TRANS-TIB ACL W/ SAWBLD |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 8/18/2023, it was reported by a sales representative via sems that an ar-1897s transtibial acl disposables kit had 2 of 5 guide pins that did not come with the beath pin in them.This was discovered upon opening the package during a procedure, with no reported adverse event or patient harm.Additional information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.The complaint device was not received for evaluation.Based on the information provided, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to a manufacturing issue; however, due to the device not being returned, we are unable to confirm the reported event.
|
| |
|