| Catalog Number |
05168791190 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for two patient samples tested with gluc3 (glucose) on a cobas 8000 c702 module.No units of measure were provided.The first sample initially resulted in a glucose value of 1.02 and it repeated as 6.48.The second sample initially resulted in a glucose value of 2.36 and it repeated as 5.48.
|
| |
|
Additional Manufacturer Narrative
|
|
The photometer check result was normal.The field service engineer performed cleaning and maintenance on the analyzer.Precision studies were performed and the results were ok.The investigation determined the issue was resolved by the field service engineer's actions.
|
| |
|