| Model Number |
232-10-85-300 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2023, a patient underwent an anterior lumbar interbody fusion procedure.After inserting the spacer, pilot hole prep was performed for the graft bolt using the trephine punch.The trephine was properly seated in the center hole and deployed.It was reported half-way during insertion the graft bolt snapped.The top portion was removed leaving the threaded end within in the spacer implanted in the patient.Lastly, the surgeon implanted posterior percutaneous pedicle screws.At the end of the procedure, it was reported the spacer was well placed.There were no patient symptoms or complications as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The top portion of the graft bolt has not returned for evaluation.The identifying lot number was not provided.The root cause could not be determined.If additional information is provided, a supplemental report will be submitted.Labeling review: "warnings/cautions/precautions: care should be taken in performing screw hole preparation to facilitate a proper graft bolt insertion trajectory and implantation.Confirm under fluoroscopy that the graft bolt insertion angle is as close as possible to a 40° trajectory.A shallow or incorrect graft bolt trajectory may result in encroachment of the spacer and lead to graft bolt breakage.".
|
| |
|