It was reported that after a reconstruction of the medial patellofemoral ligament in which a biosure anchor was used, on the (b)(6) 2023 the patient´s wound started to discharge oozing and poor healing was noted.On the (b)(6) 2023 a debridement procedure was performed, and the healing improved after the procedure.The implant remained in the patient after the debridement.The patient has been discharged.
|
|
H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.H11: h2: corrected data on b1, b2, d6b should be left in blank, the anchor was not explanted; and h6 (health effect - impact code).
|