• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER INC TORNIER PERFORM REVERSED LATER ALIZED BASEPLATE 3MM DIA 25MM; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC TORNIER PERFORM REVERSED LATER ALIZED BASEPLATE 3MM DIA 25MM; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number DWJ502
Medical Device Problem Codes Device Difficult to Setup or Prepare (1487); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
The baseplate initially couldn't accommodate the center compression screw.Efforts were made to reset the spacer, but it didn't work with the t20 driver or a long-handled screwdriver.After multiple unsuccessful attempts, an alternative baseplate was used.The reporter attempted to reset the spacer after the procedure but couldn't get the screw to fit properly.It's believed that the spacer in the baseplate was meant for a +6mm baseplate.The issue happened during a primary reverse, involving a 6.5 30 screw (dwj130), and it was ultimately resolved by switching to a different baseplate.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing or design related problems were found during the investigation.Based on the investigation, the event involved a product problem indicating a non-conformity, adverse trend, or unanticipated hazard due to a manufacturing issue which has been elevated to an nc to investigate further and determine the root cause of the event.If any additional information is provided, the investigation will be reassessed.H3 other text : device disposition is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER PERFORM REVERSED LATER ALIZED BASEPLATE 3MM DIA 25MM
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17750366
Report Number0001649390-2023-00232
Device Sequence Number18022459
Product Code PHX
UDI-Device Identifier00846832092123
UDI-Public00846832092123
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K161742
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDWJ502
Device Lot Number1756123013
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/14/2023
Initial Report FDA Received Date09/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/11/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-