• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VOLCANO CORPORATION EAGLE EYE PLATINUM ST RX DIGITAL IVUS CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VOLCANO CORPORATION EAGLE EYE PLATINUM ST RX DIGITAL IVUS CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number 85900PST
Medical Device Problem Codes Material Separation (1562); Sharp Edges (4013)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an eagle eye platinum catheter was used in a coronary procedure.At the beginning of the procedure, the catheter worked fine; however, it stopped working mid procedure.No patient injury reported.During the product return evaluation, a portion of the kapton was lifted and missing with sharp edges of non-malleable material observed.This product problem is being submitted due to missing material with sharp edges observed.There is a potential for harm if the malfunction were to recur.
 
Additional Manufacturer Narrative
This case was reviewed and investigated according to the manufacturer¿s policy.Block h6: the probable cause of the missing material likely occurred during handling, while advancing through the introducer sheath or into the y-connector.Blocks h7 & h9: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Additional Manufacturer Narrative
A: patient information were inadvertently missed.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EAGLE EYE PLATINUM ST RX DIGITAL IVUS CATHETER
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
VOLCANO CORPORATION
3721 valley centre drive #500
san diego CA 92130
Manufacturer (Section G)
VOLCARICA S.R.L.
coyol free zone &business park
b37
alajuela
CS  
Manufacturer Contact
ayse kharodawala
3721 valley centre drive #500
san diego, CA 92130
858720-4063
MDR Report Key17750695
Report Number3008363989-2023-00042
Device Sequence Number15256476
Product Code OBJ
UDI-Device Identifier00845225002312
UDI-Public(01)00845225002312(11)230609(17)250609(10)0302923846
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K120697
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/09/2025
Device Model Number85900PST
Device Catalogue Number85900PST
Device Lot Number0302923846
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/14/2023
Supplement Report FDA Received Date10/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE WIRE, MFG AND SIZE UNK; INTRODUCER SHEATH MFG AND SIZE UNK
Patient Age73 YR
Patient SexFemale
-
-