• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STOCKERT GMBH SMARTABLATE¿ SYSTEM RF GENERATOR (US); CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STOCKERT GMBH SMARTABLATE¿ SYSTEM RF GENERATOR (US); CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number M490007
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Fistula (1862)
Date of Event 08/07/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The hardware investigation has begun but it has not been completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Biosense webster inc.'s reference number (b)(4) has two reports: (1) manufacture report number # 2029046-2023-02077 for product code d134804 (thermocool® smart touch® sf bi-directional navigation catheter); (2) importer report number # 2029046-2023-50016 product code m490007 (smartablate¿ system rf generator (us)).
 
Event or Problem Description
It was reported that a patient underwent a paroxysmal atrial fibrillation (afib) ablation procedure that included thermocool® smart touch® sf bi-directional navigation catheter and smartablate¿ system rf generator (us).The patient experienced esophageal fistula that required surgical intervention.Patient was treated on (b)(6) 2023 and was in surgery (b)(6) 2023 for an esophageal fistula that was found.The patient was in stable condition at that time.The generator had been used for several other successful procedures since the case on (b)(6) 2023.Additional information was received.The physician did not provide a causality opinion for the cause of this adverse event.Patient required extended hospitalization, they were readmitted for this as it was approximately two weeks post procedure.Account is not requesting service of the generator at this time (generator has been used for dozens of cases since).The adverse event was assessed as mdr reportable under both the thermocool® smart touch® sf bi-directional navigation catheter and smartablate¿ system rf generator (us).
 
Additional Manufacturer Narrative
If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Importer reference number: (b)(4) during an internal review on 15-sep-2023, noted the following corrections to the 3500a initial: in h10.Additional manufacturer narrative reported, ¿biosense webster inc.'s reference number (b)(4) has two reports:¿ and it should have stated, ¿biosense webster inc.'s reference number (b)(4) has two reports:." in f14.Manufacturer address street line 1 reported, ¿boetzinger strasse 72¿ and it should have been processed as ¿boetzinger strasse 31¿.H6.Type of investigation was blank and it should have been processed to "type of investigation not yet determined (b21)".Investigation finding was blank and it should have been processed to "results pending completion of investigation (c21).Investigation conclusions was blank and it should have been processed to "conclusion not yet available (d16)".
 
Additional Manufacturer Narrative
It was reported that a patient underwent a paroxysmal atrial fibrillation (afib) ablation procedure that included the smartablate¿ system rf generator (us).The patient experienced esophageal fistula that required surgical intervention.The investigation was completed on 30-oct-2023.The unit was performing as intended; therefore, no failure was found with the system.A device history record evaluation was performed for the finished device number, and no internal actions related to the reported complaint condition were identified.All devices are manufactured, inspected, and released to approved specifications as part of biosense webster's quality process.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Importer reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTABLATE¿ SYSTEM RF GENERATOR (US)
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
STOCKERT GMBH
boetzinger strasse 31
freiburg, b-w D-791 11
GM  D-79111
MDR Report Key17751061
Report Number2029046-2023-50016
Device Sequence Number6664875
Product Code LPB
UDI-Device Identifier04260166371390
UDI-Public04260166371390
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/17/2023
Report Date (Section F) 08/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberM490007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/30/2023
Device Age8 YR
Event Location Hospital
Date Report to Manufacturer08/24/2023
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date09/14/2023
Supplement Report FDA Received Date09/15/2023
11/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
THMCL SMTCH SF BID, TC, F-J
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient SexUnknown
-
-