| Catalog Number |
306616 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd safetyglide¿ needle the needle was clogged.There was no report of patient impact.The following information was provided by the initial reporter: we tested another needle with water from the same lot as #1 and it happened again- we have pulled all needles from stock.I have 4 boxes + 12 i have pulled: are you able to provide the date of incident? (b)(6) 2023, (b)(6) 2023, (b)(6) 2023 are the most recent- we have the needle from today.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: a device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: no samples or photos received by our quality team for evaluation.Furthermore, a device history record review was completed for provided batch number 2188470.According to the sampling plan applied for product performance, this batch was accepted and released without defects being noted during the final assembly or visual inspections.Additional device histories were reviewed for each lot of needles used in the manufacturing of this batch.During the history review, one lot was identified as having suffered from clogged needles due to silicone.It could be possible some samples are escapes from the incident that occurred during production.Corrective and preventative actions have been implemented.Several quality initiatives have been implemented on our manufacturing line to ensure that the needle lubricant application is properly applied during the manufacturing process.Additionally, a monitoring program is also in place to verify the needle lubricant is applied uniformly to the needle.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd safetyglide¿ needle the needle was clogged.There was no report of patient impact.The following information was provided by the initial reporter: we tested another needle with water from the same lot as #1 and it happened again- we have pulled all needles from stock.I have 4 boxes + 12 i have pulled.1.Are you able to provide the date of incident? (b)(6) 2023 are the most recent- we have the needle from today.
|
| |
|