| Model Number |
WB50402W |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Reports are being submitted the foot switch and high frequency generator.Please refer to the following events: patient identifier of (b)(6) is related to model number: wb50402w, serial/lot number: unknown.Patient identifier of (b)(6) is related to model number: wb91051w, serial/lot number: (b)(6).This event is under investigation.A supplemental report will be submitted upon receiving additional information.
|
| |
|
Event or Problem Description
|
|
The customer reported that when using the foot switch, double with the high frequency generator during a procedure, an e433 error message was displayed.The customer noted the foot switch was in bad condition.There was no reported patient harm or impact due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the root cause of the event was unable to be identified.The event likely occurred due to the following: 1.A temporarily sticking reed contact.2.A temporary short circuit in the connector plug.3.A temporary short circuit due to moisture-induced leakage currents.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|