• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RANGER PACLITAXEL-COATED PTA BALLOON CATHETER; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLASTY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RANGER PACLITAXEL-COATED PTA BALLOON CATHETER; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLASTY CATHETER Back to Search Results
Model Number 1973-03
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of a ranger drug coated balloon catheter.The outer shaft, inner shaft, balloon, and tip were visually and microscopically examined.Visual and microscopic examination revealed no damages.The device was functionally tested and the balloon was able to deflate without any issues.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that slow deflation occurred.The 99% stenosed target lesion was located in the moderately tortuous and moderately calcified superficial femoral artery (sfa).A 4.0 x 150mm, 150cm ranger paclitaxel-coated pta balloon catheter was selected for use.During the procedure, the deflation was very slow.The procedure was completed with this device.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RANGER PACLITAXEL-COATED PTA BALLOON CATHETER
Common Device Name
DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLASTY CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
HEMOTEQ AG
adenauerstrasse 15
wuerselen 52146
GM   52146
Manufacturer Contact
rachel shields shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17756303
Report Number2124215-2023-50561
Device Sequence Number7094528
Product Code ONU
UDI-Device Identifier08714729975861
UDI-Public08714729975861
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P190019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/19/2024
Device Model Number1973-03
Device Catalogue Number1973-03
Device Lot Number04791H22
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/23/2023
Initial Report FDA Received Date09/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-