| Catalog Number |
Z990G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 9/15/2023.H6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: when were the sutures broke (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify for each of the 2 involved devices.Please provide the source or name of person providing answers to follow-up questions (not the person relaying/submitting answers to loc or chu.H3 investigational summary: the product was returned to ethicon for evaluation.The functional testing was conducted on the returned devices.The returned samples determined that it was received, eight unopened samples that pertain to product code z990g.In order to evaluate the condition of the returned samples, the packets were opened.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand and no issues related to breakage sutures or anomalies were observed during the evaluation.The functional test was performed using instron equipment and the tensile force was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Related reports: 2210968-2023-06767.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a c-section procedure on (b)(6) 2023 and suture was used.During the procedure, two of the sutures broke while using them during a c-section.No patient harm.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Additional information has been requested and received.If further details are received at a later date a supplemental medwatch will be sent.When were the sutures broke (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify for each of the 2 involved devices.My understanding is that both sutures broke during use on the patient but i don't know for sure and don't have additional information.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|