MAUDE Adverse Event Report: ZIMMER BIOMET, INC. JUGGERSTITCH STRAIGHT IMPLANT; SOFT-TISSUE/MESH ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. JUGGERSTITCH STRAIGHT IMPLANT; SOFT-TISSUE/MESH ANCHOR, NON-BIOABSORBABLE Back to Search Results
Catalog Number 110024772
Medical Device Problem Code Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).D10: juggerstitch curved implant cat#110024773, lot#65867720.G2: (b)(6).Visual examination of the returned product identified sample one has been cycled one time and both blue anchors remain in the needle; sample 2 has been cycled 2 times and both blue anchors have been deployed from the needle.For the device that has been cycled twice with both anchors deployed (sample 2), it was noted that a blue fiber is present on the pusher wire; however, it cannot be confirmed where the piece originated from.Unable to verify item/lot information as there are no product identifiers on the product and there was no packaging returned with the devices.However, the device that has been cycled 2 times and has the anchors deployed is a curved front-end assembly and the device that has been cycled once and has retained the blue anchors has a straight front-end assembly.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The reported event of the anchor not deploying for part 110024772 is confirmed from product return.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the anchor didn¿t deploy during a procedure.No harm or impact to the patient was reported.Attempts have been made and no additional information on the reported event is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUGGERSTITCH STRAIGHT IMPLANT
Common Device Name
SOFT-TISSUE/MESH ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17756959
Report Number0001825034-2023-02176
Device Sequence Number15145758
Product Code MBI
UDI-Device Identifier00880304690066
UDI-Public(01)00880304690066(17)270929(10)466980
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K191459
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number110024772
Device Lot Number466980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2023
Initial Report FDA Received Date09/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Patient SexUnknown
-
-