MAUDE Adverse Event Report: RIVERPOINT MEDICAL LLC BROADBAND LOOP; NONABSORBABLE SURGICAL SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RIVERPOINT MEDICAL LLC BROADBAND LOOP; NONABSORBABLE SURGICAL SUTURE Back to Search Results
Model Number CM-0324N
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/24/2023
Type of Reportable Event Serious Injury
Event or Problem Description
According to the reporter, "it was reported that the patient underwent a primary reverse shoulder replacement on (b)(6) 2023.Subsequently, the patient underwent a revision surgery on (b)(6) 2023 to exchange the humeral bearing, glenosphere, and taper adapter.There was a continuation of the event where there was continuous wound ooze, so all implants were removed on (b)(6) 2023 to exchange the humeral bearing, glenosphere, and taper adapter.There was a continuation of the event where there was continuous wound ooze, so all implants were removed on (b)(6) 2023.".
 
Additional Manufacturer Narrative
The device was discarded and therefore was not returned.A review of the device history record was performed and all product met requirements prior to release.Investigation is still ongoing.No cause for the event has been established at this time.This report and use of categorical definitions required by fda 3500a does not constitute an admission by riverfront medical or its employees that riverpoint medical or its employees have caused or contributed to the event described in this report.Riverpoint medical has filed this information to comply with the medical device reporting regulation 21 cfr 803.If additional information is provided to riverpoint medical regarding this event, a supplementary 3500a form will be submitted as required by the fda.
 
Additional Manufacturer Narrative
Review of sterilization records showed that all products met sterilization requirements prior to release.Retain samples were pulled and underwent third-party sterility testing.The samples tested were found to be negative for growth after 14 days incubation.A cause for the event cannot be established.This report and use of categorical definitions required by fda 3500a does not constitute an admission by riverpoint medical or its employees that riverpoint medical or its employees have caused or contributed to the event described in this report.Riverpoint medical has filed this information to comply with the medical device reporting regulation 21 cfr 803.If additional information is provided to riverpoint medical regarding this event, a supplementary 3500a form will be submitted as required by the fda.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROADBAND LOOP
Common Device Name
NONABSORBABLE SURGICAL SUTURE
Manufacturer (Section D)
RIVERPOINT MEDICAL LLC
825 ne 25th ave
portland OR 97232
Manufacturer (Section G)
RIVERPOINT MEDICAL LLC
825 ne 25th ave
portland OR 97232
Manufacturer Contact
edwin anderson
825 ne 25th ave
portland, OR 97232
MDR Report Key17756964
Report Number3006981798-2023-00056
Device Sequence Number15145760
Product Code GAT
UDI-Device Identifier00810020081244
UDI-Public00810020081244
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K190817
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model NumberCM-0324N
Device Lot Number21081224
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2023
Supplement Date Received by Manufacturer08/17/2023
Initial Report FDA Received Date09/15/2023
Supplement Report FDA Received Date11/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-