| Model Number |
3CX*FX15RW30C |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during set up venous inlet temp probe was broken.No patient involvement.Product was changed out.Procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.Component code: 918- probe.Health effect ¿ impact code: 2645- no patient involvement.Health effect ¿ clinical code: 582- no clinical signs, symptoms or conditions.Medical device problem code: 1069 - break.Investigation findings: 3233 - results pending completion of investigation.Investigation conclusions: 11 - conclusion not yet available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date); g3 (date received by manufacturer) ; g6 (indication that this is a follow-up report) ; h2 (follow-up due to additional information and device evaluation); h3 (device evaluated by manufacturer) ; h4 (device manufacture date); h6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation findings: 3259- improper physical structure.Investigation conclusions: 4307 - cause traced to component failure.The affected sample was inspected upon receipt and confirmed that the venous temp probe was broken.A representative retention sample was inspected for damage with no damage noted on the device.All capiox units are 100% visually inspected at several points in the production and packaging processes.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
| |
|
Event or Problem Description
|
|
Broken venous inlet temp probe.
|
| |
|
Event or Problem Description
|
|
Broken venous inlet temp probe.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on (b)(6) 2023. upon further investigation of the reported event, the following information is new and/or changed: b3 (updated date of event).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information and correction). a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11. all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
| |
|