| Catalog Number |
67PFSS40 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient was decompensating and needed a trach exchange.A new trach tube was opened.Pre-testing inflation of the cuff was completed to ensure it was not compromised.The cuff, however, only inflated on one side.Another trach package was opened.Subsequent check ascertained effectiveness of new trach.Adverse effects have not been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: additional information is provided in h.2., and h.6.No product or photographic evidence were provided to aid in this investigation.The complaint report offered insufficient details to determine whether this product functioned as intended or was used in a manner consistent with its instructions for use (ifu) or failed to meet product specifications.Lacking any additional evidence, this complaint has been closed as unconfirmed.If the product is returned, icu medical will reopen this complaint for further investigation.A device history (dhr) review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.For all enquiries or follow-up questions related to the record, do not use regulatory.Responses@smiths-medical.Com located in sections g.1., please direct those to the following: regulatory.Responses@icumed.Com.
|
| |
|
Search Alerts/Recalls
|