| Model Number |
CYF-5A |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cystonephrofiberscope had a damaged forceps port.This was found when cleaning up after a diagnostic medical examination.The procedure had been completed prior to the event with no delay, using the same device without any problems.There was no report of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation and the reported event was confirmed; the forceps channel port was found deformed.Other evaluation findings are as follows: the bending angle in the up direction did not meet the standard value due to wear of the angle wire; the light guide bundle was slipping down; the control unit was sticky due to water leakage; the upward/downward plate was also sticky; the image guide bundle was stained; and a foggy image occurred due to damage on the objective lens.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over three years since the subject device was manufactured.Based on the results of the investigation, the root cause of the damaged forceps port could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to correct the legal manufacturer¿s contact information and facility registration number.The facility registration number is 3002808148.
|
| |
|