| Model Number |
N/A |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the distributorship that the products were found to be nonconforming.Incoming inspection team found debris in the sterile package.No reported patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign: japan.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.Visual examination of the returned product found (1) sealed cavity with one piece of loose debris exceeding the acceptable limits.This complaint has been confirmed by evaluation of the returned product.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The condition of the device when it left zimmer biomet is considered non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|