| Model Number |
35615 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the ro marker moved along the catheter shaft.This accustick ii introducer catheter was selected for use to treat biliary duct obstruction.During percutaneous insertion, the ro marker slid back along the sheath proximally toward the physician's hands.It was confirmed that the ro marker didn't detach/separate from the tip of the device (only slid back).The procedure was completed with another of the same device.No complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of media: this accustick ii catheter was disposed at the facility and not available for analysis; however, the physician provided a photo for review.Review of the photo revealed that the ro marker had moved out of the intended location in the distal section.
|
| |
|