| Catalog Number |
EV08262CD |
| Medical Device Problem Codes |
Difficult to Insert (1316); Device Dislodged or Dislocated (2923); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the external iliac artery via the common femoral artery, the stent allegedly got caught in valve of sheath when tried to advance.It was further reported that the stent was not very tight on the balloon either and slid off way too easy.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 10/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.One photo was provided and reviewed.The photo shows the valeo balloon deflated and bloody.The stent is seen deformed and dislodged from the balloon.No other anomalies noted.Therefore, the investigation is confirmed for the reported stent dislodgement and identified stent deformation as the stent is seen deformed and dislodged from the balloon.However, the investigation is inconclusive for the reported sheath insertion difficulty as functional testing for the same could not be performed due to the condition of the device.A definitive root cause for the alleged sheath insertion difficulty, stent dislodgement and identified stent deformation could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 10/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the external iliac artery via the common femoral artery, the stent allegedly got caught in valve of sheath when tried to advance.It was further reported that the stent was not very tight on the balloon either and slid off way too easy.The procedure was completed using another device.There was no reported patient injury.
|
| |
|