| Model Number |
CD3231-40 |
| Medical Device Problem Codes |
High impedance (1291); Loose or Intermittent Connection (1371); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient presented to clinic for follow up.Device interrogation revealed high pacing impedance on the right ventricular (rv) lead.The physician considers high pacing impedance is due to inadequate insertion on the implantable cardioverter defibrillator (icd) header.During repositioning, the set screw was unable to be tightened.The physician elected to explant and replace the icd.The patient condition was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported field event of lead connection issue was confirmed.Septum material was observed inside the right ventricular lead set screws hex cavity.This did not allow the set screw to be tightened and secured properly in the field, resulting in high lead impedance.The issue was consistent with having occurred during the procedure.The device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
|
| |
|