| Medical Device Problem Codes |
Mechanical Problem (1384); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation of the reported event is in process.A follow-up mdr will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The user facility reported that during a patient procedure, their 5095 surgical table "stopped working".No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Through follow up, the user facility stated that the procedure was completed successfully.The user facility also stated that the table would not respond to hand control commands.The user facility utilized the back-up hand controls to complete the procedure.The 5095 surgical table subject of the reported event was returned to steris for evaluation.The evaluation determined that the table had torn bellows and a bent bellow frame.As the table had a bent bellow frame, the table was unable to lower to its lowest height.The customer was provided with a replacement table.No additional issues have been reported.
|
| |
|