| Model Number |
M00545180 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 44 was used in the ampulla during an endoscopic retrograde cholangiopancreatography (ercp) procedure.The exact procedure date was unknown.During the procedure, while cutting, the cutting wire of the autotome rx 44 broke.The working length was also twisted, which resulted to a wrong orientation.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: imdrf device code a051201 captures the reportable event of cutting wire anchor dislodged.
|
| |
|
Search Alerts/Recalls
|