| Catalog Number |
SRLP-360 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
07/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported, capsular contracture, baker grade iii.This record is for the right side.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication.And analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: capsular contracture, baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported capsular contracture baker grade iii.This record is for the right side.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the device analysis grid, the assessments of the complaint are: ¿ capsular contracture: unable to observe since it is not related to the device.Additional observations: ¿ crease fold and wear abrasion observed on the device.¿ brown particles observed on the device.No further actions are required as the device was implanted.
|
| |
|
Search Alerts/Recalls
|