| Model Number |
VTICM5 13.7 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/18/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information: h6-investigation type 4110: lens work order search- no similar complaint event(s) within associated lots were found.Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 13.7mm vticm5 13.7 implantable collamer lens of -10.00/2.0/120 (sphere/cylinder/axis) was implanted into the patients right eye (od) on an unknown date.Excessive vault was observed.The lens was explanted and a replacement lens was implanted of shorter length.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Claim#: (b)(4).
|
| |
|
Search Alerts/Recalls
|