| Catalog Number |
MQK1210 |
| Medical Device Problem Codes |
Nonstandard Device (1420); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: the manufacturing location was selected as unknown due to system limitations.The manufacturing location for this product is bd-santo domingo.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 05/2026).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during a breast biopsy procedure, the needle allegedly split into two full pieces, when the needle was removed from the breast.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during a breast biopsy procedure, the needle split into two full pieces, when the needle was removed from the breast.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.One electronic photo was provided and reviewed.In that photo, one marquee biopsy device was shown in which the cannula part of the needle was detached from the marquee device.Based on the photo review, the reported detachment of device or device component can be confirmed.Therefore the investigation is confirmed for the reported detachment of device or device component as the photo shows the detachment of needle from the device.A definitive root cause for the reported detachment of device or device component could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiry date: 05/2026), g3, h6 (method).H11: g1, h6 (result, conclusion).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.One electronic photo was provided and reviewed.In that photo, one marquee biopsy device was shown in which the cannula part of the needle was detached from the marquee device.It is noted there is no flare to the cannula end.Based on the photo review, the reported detachment of device or device component and identified nonstandard device can be confirmed.Therefore the investigation is confirmed for the reported detachment of device or device component as the photo shows the detachment of needle from the device.The investigation is confirmed for the nonstandard device issue as the cannula end was found to be not flared.The root cause for the reported detachment issue is likely due to the condition of the cannula (e.G., nonhomogeneous, and not flared to acceptable criteria).The root cause for the identified nonstandard device issue is determined to be manufacturing related.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiry date: 05/2026), g3, h6 (device, result).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a breast biopsy procedure, the needle split into two full pieces, when the needle was removed from the breast.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during a breast biopsy procedure, the needle split into two full pieces, when the needle was removed from the breast.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.One electronic photo was provided and reviewed.In that photo, one marquee biopsy device was shown in which the cannula part of the needle was detached from the marquee device.It is noted there is no flare to the cannula end.Based on the photo review, the reported detachment of device or device component and identified nonstandard device can be confirmed.Therefore the investigation is confirmed for the reported detachment of device or device component as the photo shows the detachment of needle from the device.The investigation is confirmed for the identified nonstandard device issue as the cannula end was found to be not flared.The root cause for the reported detachment issue is likely due to the condition of the cannula (e.G., nonhomogeneous, and not flared to acceptable criteria).The root cause for the identified nonstandard device issue is determined to be manufacturing related.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|