MAUDE Adverse Event Report: BIOMET 3I 3I T3¿ WITH DCD¿ EX HEX TAPERED IMPLANT 4 X 11.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I 3I T3¿ WITH DCD¿ EX HEX TAPERED IMPLANT 4 X 11.5MM; DENTAL IMPLANT Back to Search Results
Catalog Number BNET411
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Osteolysis (2377)
Date of Event 01/17/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The doctor reports that the dental implants located in dental position number #32, #42 and #43 failed due to peri-implantitis.The procedure was not concluded by placing another implants.Sites presented inflammation and edema.Patient felt pain.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number: (b)(4).A4: patient weight unknown / not provided.D10.Concomitant medical products bnet511, 3i t3 with dcd ex hex tapered implant 5 x 11.5mm lot: 2021040231 and oss313, osseotite implant 3.75 x 13mm lot 2022030168.H6: event problem codes ¿ patient code: 1932 inflammation and 4577 edema.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3¿ WITH DCD¿ EX HEX TAPERED IMPLANT 4 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key17771538
Report Number0001038806-2023-01757
Device Sequence Number20627248
Product Code DZE
UDI-Device Identifier00844868033066
UDI-Public(01)00844868033066(17)260331(10)2021030485
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K133049
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date03/30/2026
Device Catalogue NumberBNET411
Device Lot Number2021030485
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DENTAL IMPLANT.
Patient Age67 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-