EXACTECH, INC. LOGIC CR FEMORAL CEM, RIGHT, SZ 5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, CEMENTED, POLYMER/METAL/POLYMER Back to Search Results
Model Number N/A
Medical Device Problem Codes Mechanical Problem (1384); Insufficient Device Problem Information (3190); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Synovitis (2094); Ankylosis (4527); Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 10/14/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending investigation.
 
Event or Problem Description
As reported via legal documentation the patient had a right knee replacement on (b)(6) 2013.Approximately 2 years and 8 months after the initial procedure the patient had a right knee revision on (b)(6) 2015.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.The patient has filed a short-form complaint in a multidistrict litigation titled in re: exactech polyethylene orthopedic products liability litigation, mdl no.3044, and pending in the eastern district of new york.Per the transfer order creating this multidistrict litigation, ¿plaintiffs allege that their knee or hip replacement devices¿failed prematurely because of degradation of the device¿s polyethylene component, which resulted in the premature removal (or planned removal) of the prosthesis at issue.¿ because the patient has filed a suit in this multidistrict litigation, the patient appears to allege that the patient was injured as a result of premature wear of an exactech polyethylene device.
 
Additional Manufacturer Narrative
D10.Concomitants: 02-012-49-5009 - logic cr tib insert slope++, sz 5, 9mm 2298729.02-010-03-0350 - logic cr femoral cem, right, sz 5 2080781.02-012-45-5040 - lgc tibial fit tray cem sz 5f / 4t 2539278.200-02-38 - three peg patella 38mm 2251683.Further information and or investigation results will be provided within 30 days of receipt.
 
Event or Problem Description
As reported via legal documentation the patient had a right knee replacement on (b)(6) 2013 due to end stage osteoarthritis.Approximately 2 years and 8 months after the initial procedure the patient had a right knee revision on (b)(6) 2015.Revision op report postoperative diagnosis: fibrous ankylosis, right knee, with loose femoral component.The surgeon noted that after resection of the posterior cruciate and notch, the component appeared to be intact, but on gentle manipulation it easily loosened removing the metal from the overlying cement that stayed on the bone.The tibial base was still intact and left alone.There is no other patient demographic or medical history available.There is no device return.There are no photos or other images of the device provided.No additional information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, femoral loosening, and loss of range of motion, or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC CR FEMORAL CEM, RIGHT, SZ 5
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, CEMENTED, POLYMER/METAL/POLYMER
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key17772085
Report Number1038671-2023-02303
Device Sequence Number17932830
Product Code JWH
UDI-Device Identifier10885862158147
UDI-Public10885862158147
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K111400
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/16/2021
Device Model NumberN/A
Device Catalogue Number02-010-03-0350
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2023
Supplement Date Received by Manufacturer09/13/2023
12/30/2024
Initial Report FDA Received Date09/19/2023
Supplement Report FDA Received Date10/31/2023
01/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/17/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age59 YR
Patient SexMale
-
-