| Catalog Number |
600-9008-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023, abbott point of care (apoc) was contacted by a customer regarding i-stat chem8+ cartridge that yielded a discrepant potassium result of >9 mmol/l on a female patient.There was no patient information at the time of this report.Method result sample.I-stat >9.0 mmol/l a.Antech 4.3 meq/l a.Test/collection times not avail.Product lot# information not provided but requested.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggested that product was not performing within the variability.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 31-oct-2023.A review of the device history records (dhrs) confirmed the cartridge lots met finished goods release criteria.Retained cartridge testing met the acceptance criteria found in q04.01.003 rev.Al, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been determined for chem8+ cartridge lots h23210, h23214 or h23237.
|
| |
|