MAUDE Adverse Event Report: ABBOTT POINT OF CARE INC. I-STAT CHEM8+ CARTRIDGE; CHEM8+ CARTRIDGE,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE INC. I-STAT CHEM8+ CARTRIDGE; CHEM8+ CARTRIDGE, Back to Search Results
Catalog Number 600-9008-25
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2023, abbott point of care (apoc) was contacted by a customer regarding i-stat chem8+ cartridge that yielded a discrepant potassium result of >9 mmol/l on a female patient.There was no patient information at the time of this report.Method result sample.I-stat >9.0 mmol/l a.Antech 4.3 meq/l a.Test/collection times not avail.Product lot# information not provided but requested.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggested that product was not performing within the variability.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident: # (b)(4).The investigation was completed on 31-oct-2023.A review of the device history records (dhrs) confirmed the cartridge lots met finished goods release criteria.Retained cartridge testing met the acceptance criteria found in q04.01.003 rev.Al, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been determined for chem8+ cartridge lots h23210, h23214 or h23237.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT CHEM8+ CARTRIDGE
Common Device Name
CHEM8+ CARTRIDGE,
Manufacturer (Section D)
ABBOTT POINT OF CARE INC.
400 college road
princeton NJ 08540
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
ottawa, ontario K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
185 corkstown road
princeton, NJ K2H 8-V4
CA   K2H 8V4
6133123525
MDR Report Key17773148
Report Number2245578-2023-00128
Device Sequence Number7022674
Product Code JGS
UDI-Device Identifier10054749005519
UDI-Public10054749005519
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K053110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number600-9008-25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/01/2023
Supplement Date Received by Manufacturer10/31/2023
Initial Report FDA Received Date09/19/2023
Supplement Report FDA Received Date11/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-