| Catalog Number |
328431 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd insulin syringes with bd ultra-fine¿ needle the hub separated from the device.There was no report of patient impact.The following information was provided by the initial reporter: consumer reported found 2 syringes when removed needle shield the hub assembly stayed in the shield.Lot # 0160994.Catalog# 328431.Date of event unknown this box.Sample status awaiting sample 1 out of the 2 samples.
|
| |
|
Additional Manufacturer Narrative
|
|
B.3.Date of event is unknown; awareness date has been used for this field.H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 05-oct-2023.H.6.Investigation summary: samples were received and an investigation was performed.This is the 3rd complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd insulin syringes with bd ultra-fine¿ needle the hub separated from the device.There was no report of patient impact.The following information was provided by the initial reporter: consumer reported found 2 syringes when removed needle shield the hub assembly stayed in the shield lot # 0160994;.Catalog# 328431; date of event unknown this box ; sample status awaiting sample 1 out of the 2 samples.
|
| |
|