• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SELF CENT HIP 52X28 BLU; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SELF CENT HIP 52X28 BLU; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI Back to Search Results
Catalog Number 103552000
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Fall (1848); Joint Dislocation (2374); Joint Laxity (4526)
Date of Event 09/06/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # : pc-(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Patient slipped in shower at home and fell (b)(6) 2023.Because of the fall the patient's incurred a traumatic hip dislocation.Attending to emergency room physician was able to relocate the hip to the acetabulum, but upon post reduction radiograph it appeared the bipolar shell had disassociated from the 28mm ball which was still on stem.Bipolar shell was floating around the inner capsule area, creating an unstable hip construct.Patient was admitted as inpatient on (b)(6) 2023 and surgery open reduction/hip revision scheduled for (b)(6) 2023.Revision surgery revealed that the 28mm ball disassociated from the poly liner, but the poly liner did not dissociate from the outer metal shell.A new 28mm of the same neck length and self-centering bipolar shell was presented and implanted.The hip was then reduced without incident and was found to be stable.Original surgeon would like for implants to be reviewed.Doi: (b)(6) 2023.Dor: (b)(6) 2023.Affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
 
Event or Problem Description
Additional information was received and stated that according to the surgeon, the disassociation came from the er physician trying to relocate the hip.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF CENT HIP 52X28 BLU
Common Device Name
HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17775750
Report Number1818910-2023-19077
Device Sequence Number20770935
Product Code KWY
UDI-Device Identifier10603295003427
UDI-Public10603295003427
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K812672
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number103552000
Device Lot NumberJR4055
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2023
Supplement Date Received by Manufacturer09/19/2023
10/10/2023
10/13/2023
Initial Report FDA Received Date09/19/2023
Supplement Report FDA Received Date09/21/2023
10/10/2023
10/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICUL/EZE BALL 28 +1.5 GR
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexMale
-
-