| Model Number |
401932 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
07/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
During a pulmonary vein isolation procedure, while attempting to position the catheter within the heart, the catheter became lodged inside the eustachian ridge.Right internal jugular vein access was obtained and a non-abbott catheter (deflectable st-sf f-curve biosense webster) was used to snare the loop of the catheter from above.The catheter was removed successfully via the femoral vein and a non-abbott catheter (biosense webster cs ez steer) to complete the procedure with no further adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|