C.R. BARD, INC. (COVINGTON) -1018233 BARD® EXTENSION CABLE Back to Search Results
Catalog Number 153622
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the extension cord of the temperature sensing catheter was broken.Stated that the bladder temperature was unable to be measured due to the disconnection of lead wire.
 
Event or Problem Description
It was reported that the extension cord of the temperature sensing catheter was broken.It was stated that the bladder temperature was unable to be measured due to the disconnection of lead wire.
 
Additional Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.The device met relevant specifications.Received 1 extension cord.Visual inspection noted using a multimeter, the cord was tested for and found to have continuity.The cord was connected to an in-house temperature sensing catheter submerged in a water bath (119118 lot ngevz327) the cord was then attached to a kilo device and the temperature displayed 37.3 celsius, similar to the thermometer reading.The photo sample was returned and could not be evaluated for the reported failure.The sample meets the specification "plug and jack must be firmly secured to wire." a dhr is not required since the reported failure was unconfirmed.As the reported event is unconfirmed a labeling review is not required.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® EXTENSION CABLE
Common Device Name
BARD® EXTENSION CABLE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17780101
Report Number1018233-2023-06900
Device Sequence Number6576306
Product Code EZL
UDI-Device Identifier00801741045875
UDI-Public(01)00801741045875
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number153622
Device Lot NumberBMHP0001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2023
Supplement Date Received by Manufacturer12/05/2023
Initial Report FDA Received Date09/20/2023
Supplement Report FDA Received Date12/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-