| Catalog Number |
54-25677S |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Wound Infection (2447); Implant Pain (4561)
|
| Date of Event |
08/17/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient required revision surgery due to breakage of the variax plate and infection from the resulting open wound.Breakage of the device was confirmed on x-ray following progressive pain.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, since the device was returned and matches the failure mode reported.The device inspection revealed the following: the plate was returned and is indeed broken in two parts.The analysis of the breakage surfaces shows a fatigue pattern apparent by the shiny surfaces that appear on both sides.This is an indication that the device got broken gradually and that the two sides rubbed against each other for a while before breaking completely.Based on investigation, the root cause was attributed to a fatigue related issue.The exact root cause of what lead to the fatigue could not be determined since more information like the x-rays, patient activity levels and details is required.Some indication of the patient activity levels such as "he used his left hand on the toilet and hear it snap.He didn't fall on it." was given and could lead to believe that an important amount of force was used on the implant sporadically by the patient, but more information is required.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient required revision surgery due to breakage of the variax plate and infection from the resulting open wound.Breakage of the device was confirmed on x-ray following progressive pain.
|
| |
|