Model Number 7211
Medical Device Problem Codes Break (1069); Material Rupture (1546); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that balloon rupture and shaft break occurred requiring additional intervention.The stenosed target lesion was located in the moderately calcified right coronary artery.A non-boston scientific (bsc) stent was placed and a 4.50mm x 8mm nc emerge balloon catheter was used for post dilatation.However, during first inflation, the balloon got stuck inside of the stent when tried to deflate and when the balloon was re-inflated at increased atmospheric pressure, the balloon ruptured.Subsequently, when the system was removed, the device was noted to be fractured.The physician was able to retrieve the broken fragments with a snare.The procedure was completed with a different device.There were no patient complications reported, and the patient was discharged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17780478
Report Number2124215-2023-48261
Device Sequence Number5889585
Product Code LOX
UDI-Device Identifier08714729846314
UDI-Public08714729846314
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K141236
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/10/2025
Device Model Number7211
Device Catalogue Number7211
Device Lot Number0031396147
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/30/2023
Initial Report FDA Received Date09/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-