|
This report is being submitted to report additional information.The following sections are being reported: d4: expiration date and unique identifier (udi) number h2: if follow-up, what type h3: device evaluated by manufacturer h4: device manufacturer date h6: type of investigation h6: investigation findings h6: investigation conclusions h10: additional narratives/data zimvie did not receive one abutment back for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation is parafunctional habits/patient factors.Therefore, based on the available information, a device malfunction could not be verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|