MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION REVACLEAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION REVACLEAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 114745M
Medical Device Problem Codes Fluid/Blood Leak (1250); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported an external saline solution leak was observed originating from the arterial port of a revaclear 300 set during prime.Upon further inspection, the port was noted to be ¿fractured¿.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
H10: the device was received for evaluation in an open package.During visual inspection, the hansen port was observed damaged on the side of the data matrix code.The reported condition was verified.There are process controls in place to test the material characteristics of the dialyzer housings, which is called the hansen break test.The procedure applies to all housing parts produced in the injection molding department of the facility.A dialyzer with a missing port would not make it through the manufacturing process to final packaging.The cause of the condition could not be determined; however, the most probable cause would occur during the transportation process.Dialyzer cases that are roughly handled during shipment and travel, as well as cases that are not secured adequately during transport, or are subject to undue force, can potentially become susceptible to breakage and housing damage.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVACLEAR
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - OPELIKA
1101 jeter ave
opelika AL 36801
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17782440
Report Number3006552611-2023-00045
Device Sequence Number12234482
Product Code KDI
UDI-Device Identifier07332414126193
UDI-Public(01)07332414126193
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K130039
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 11/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/08/2026
Device Catalogue Number114745M
Device Lot NumberC623306107
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2023
Supplement Date Received by Manufacturer10/16/2023
Initial Report FDA Received Date09/20/2023
Supplement Report FDA Received Date11/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-