| Catalog Number |
2H8537 |
| Medical Device Problem Code |
Incomplete or Inadequate Connection (4037)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: additional initial reporter phone no.: (b)(6).G1: device manufacturer address 1: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a clearlink system non-dehp continu-flo solution set was unable to connect as the connector was unable to retract.This was discovered during set-up and preparation.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the device was received for evaluation.A visual inspection was performed which observed the collar was bonded to the male luer which affected the mobility or spin of the collar.The reported condition was verified.The cause of the condition was improper solvent application during the automated assembly of the device's tail end segment during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|