Catalog Number 36-2018
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer alleged that during a decompression posterior lumbar with fusion extension to l1 and anterior lumbar interbody fusion with vascular assist, one side of the jaws broke off entirely.The piece fell into the patient and was retrieved with no further harm to the patient.
 
Additional Manufacturer Narrative
The complaint was confirmed via pictures provided by the customer.However, the picture provided did not show any of the markings on the device to determine the lot number of the device.The device was discarded after the pictures were taken.Therefore, the device cannot be returned and no more pictures can be taken.Without further information, a true root cannot be determined.The complaint has been logged in the complaint system for tracking and trending.Based on the above information, no further actions are required and this can be seen as the final report.If additional information is obtained that alleges any additional patient involvement or additional information pertinent to the investigation, a subsequent follow up report will be submitted.There has been a total of (b)(4) sold of all lots with no additional complaints recorded for the jaw breaking off.Therefore, this is an isolated event.Received via user facility report# (b)(4).
 
Additional Manufacturer Narrative
The customer notified symmetry on 10/09/2023 that the device was found and available for returned.Upon receiving the device, it was evaluated and confirmed to have a manufactured date of 1410 ( oct 2014).The device showed significant signs of wear and tear and a dard spot on the break that indicitive that the jaw was cracked for sometime before the jaw broke off during the procedure.The crack could have occurred during another case, or it could have been mishandled during cleaning.Based on the product review and the risk assessment no further actions are necessary.Based on the above information, no further actions are required and this can be seen as the final report.If additional information is obtained that alleges any additional patient involvement or additional informaiton pertinent to the investigation, a subsequent follow up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMMETRY
Common Device Name
NEEDLE HOLDER
Manufacturer (Section D)
SYMMETRY SURGICAL INC
3034 owen drive
antioch TN 37013
Manufacturer (Section G)
SYMMETRY SURGICAL INC
3034 owen drive
antioch TN 37013
Manufacturer Contact
brandi meath
3034 owen drive
antioch, TN 37013
6159645290
MDR Report Key17785551
Report Number3007208013-2023-00044
Device Sequence Number20470889
Product Code HXK
UDI-Device Identifier00887482009727
UDI-Public00887482009727
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number36-2018
Device Lot Number1410
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2023
Supplement Date Received by Manufacturer10/09/2023
Initial Report FDA Received Date09/20/2023
Supplement Report FDA Received Date10/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-