| Model Number |
554-3003-001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
We apologize for the delay in filing this report.The complaint record was inadvertently misdated, resulting in a target reporting date beyond the 30 day due date.
|
| |
|
Event or Problem Description
|
|
During a procedure with the tenex system, the microtip needle separated from the rest of the hand piece.There were no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up report #01.Corrections were made to the device brand name, model number and labeling information.The device was returned and evaluated.It was confirmed that the microtip needle had separated from the rest of the handpiece.Due to the state in which the device was received, functional testing could not be performed.The fracture site did not immediately indicate a specific cause for the failure.Disassembly of the device did not reveal any further anomalies or defects.
|
| |
|